• 1. Department of Biomedical Engineering, Chinese PLA General Hospital, Beijing 100853, China;
  • 2. Department of Biomedical Engineering, Hai Nan Fen Yuan of Chinese PLA General Hospital, Sanya 572014, China;
WANGWeidong, Email: wangwd301@126.com
Export PDF Favorites Scan Get Citation

To evaluate the influence of different brands of syringes on performance parameters and application quality of syringe pump, we carried out matching experiment of infusion/syringe pump analyzer Rigel Multi-Flo, made by ALK Co. in Sweden, for 3 different types of syringes to measure the flow rate, occlusion alarm pressure, occlusion time and bolus volume of 26 syringe pumps. Data of quality control were analyzed with statistical method. The results demonstrated that there were significant differences in the flow rate, occlusion alarm pressure, occlusion time and bolus volume (P<0.01) between those of common syringes and original syringes. This study points out that the health care providers should select compatible syringes for syringe pumps, which provides evidence and guidance to assure the reasonable and safe application of syringe pumps in clinical practice.

Citation: LIUHongyun, HUANGHao, WANGWeidong. Research on Influence of Different Syringe Types on Quality Control of Syringe Pump. Journal of Biomedical Engineering, 2014, 31(6): 1316-1318,1324. doi: 10.7507/1001-5515.20140249 Copy

  • Previous Article

    CK2β Promotes Pink1/Parkin-mediated Miro1 Degradation
  • Next Article

    Research on Expression of Somatomedin B Domain of Proteoglycan 4 and Recombinant Protein Aggregation