Objective To improve health care quality and safety by monitoring the performance of PhilipsPrecedence Dual-head single photon emission computed tomography (SPECT). Methods With our own homemade point source and the center of rotation model, in accordance with NEMA standards and manufacturers' design conditions, these three indicators including energy peak position of the instrument, intrinsic uniformity and center of rotation were routinely tested between 2008 and 2012 for the Philips-Precedence Dual-head SPECT in our hospital. In addition, the quality control was performed twice a week, and every year the total number of quality control was basically the same. We calculated the results by the weighted average method. Results The 5-year average energy peaks of detector 1 and 2 were (139.23±0.32) and (138.97±0.45) keV, respectively, and they were both within the range of reference values [(140±3) keV]. In addition, the results of center of rotation were also in the normal range, and kept stable. Based on the analysis of quality-control data, for detector 1, compared with the data of 2008, there was no significant diTherence for central field of vision (CFOV) and useful field of vision (UFOV) in these three years from 2009 to 2011 (P>0.05). The diTherence was only significant between data of 2008 and that of 2012 (P<0.05). For detector 2, compared with the data of 2008, there was no significant diTherence for CFOV and UFOV in 2009 and 2011(P>0.05), while there was significant diTherent in 2010 and 2012 (P<0.05). Conclusion The uniformity of SPECT will gradually deteriorate with prolonged use. However, regular quality control and maintaining can keep the function stabilization, and enhance the availability rate.

Citation: SHENGuo-hua, ZHOULü-yi, JIAZhi-yun, YANGXiao-chuan, DENGHou-fu. Long-term Quality Control of Philips-Precedence Dual-head Single Photon Emission Computed Tomography for Five Years. West China Medical Journal, 2015, 30(10): 1906-1909. doi: 10.7507/1002-0179.20150545 Copy

  • Previous Article

    The Image Quality Analysis and Control of Whole Body Tumor Imaging with 18F-uorodeoxyglucose Positron Emission Tomography/Computer Tomography
  • Next Article

    Study on the Feasibility of Brain CT Perfusion with 70 kV Tube Voltage