• Department of Bone and Joint Surgery, the Affiliated Hospital of Southwest Medical University, Luzhou Sichuan, 646000, P.R.China;
YANG Yunkang, Email: droyyk@163.com
Export PDF Favorites Scan Get Citation

ObjectiveTo report a new type of anatomical locking plate for sternocalvicular joint, and investigate its effectiveness in treatment of sternoclavicular joint fracture and dislocation.MethodsA new type of anatomical locking plate for sternoclavicular joint was developed, which accorded with the anatomical features and biomechanical characteristics of Chinese sternoclavicular joint. By adopting the method of clinical randomized controlled study, 32 patients with the sternoclavicular joint fracture and dislocation who met the selection criteria between June 2008 and May 2015 were randomly divided into groups A and B (n=16), and the patients were treated with new anatomic locking plate and distal radial T locking plate internal fixation, respectively. There was no significant difference between 2 groups in gender, age, injured side, body mass index, cause of injury, type of injury, the time from injury to operation, and preoperative Rockwood grading score (P>0.05). The operation time, intraoperative blood loss, incision length, hospitalization time, and postoperative complications in 2 groups were recorded, and the effectiveness was evaluated by Rockwood grading score.ResultsThe operations of 2 groups completed successfully. The operation time, intraoperative blood loss, and hospitalization time in group A were significantly less than those in group B (P<0.05), but there was no significant difference in the incision length between 2 groups (t=0.672, P=0.507). All the patients were followed up 18-30 months (mean, 24 months). In group A, there were 1 case of sternoclavicular joint pain and 2 cases of wound infection; in group B, there were 1 case of sternoclavicular joint pain, 1 case of internal fixation loosening, and 1 case of sternoclavicular joint re-dislocation; there was no significant difference in complication incidence between 2 groups (P=1.000). The Rockwood grading scores at each time point after operation in 2 groups were significantly higher than those before operation. At 1 month after operation, the Rockwood grading score in group A was significantly higher than that in group B (t=2.270, P=0.031); but there was no significant difference in the Rockwood grading scores between the 2 groups at 6 months and at last follow-up (P>0.05). At last follow-up, according to the Rockwood scoring standard, the results of group A were excellent in 13 cases, good in 2 cases, poor in 1 case, the excellent and good rate was 93.75%; the results of group B were excellent in 11 cases, good in 4 cases, poor in 1 case, and the excellent and good rate was 93.75%; there was no significant difference between 2 groups (Z=–0.748, P=0.455).ConclusionThe new type of anatomic locking plate accords with the Chinese anatomical characteristics. It has the advantages of easy operative procedure, less surgical trauma, shorter operation time, less intraoperative blood loss, shorter hospitalization time, and it can achieve better results in the treatment of sternoclavicular joint fracture and dislocation.

Citation: SUN Yuanlin, YANG Yunkang, GE Jianhua, YANG Kun, XIANG Feifan, ZHOU Ju, LIANG Jie. Development and clinical application of a new type of anatomical locking plate for sternoclavicular joint fracture and dislocation. Chinese Journal of Reparative and Reconstructive Surgery, 2018, 32(3): 316-321. doi: 10.7507/1002-1892.201707081 Copy

  • Previous Article

    Biomechanics test of fixation of star-shaped six-part patellar fractures with petal-shaped poly-axial locking plate
  • Next Article

    Effectiveness comparison between minimally invasive surgery and traditional open reduction internal fixation in treatment of unstable distal radial fractures