• Department of Oncology, the Affiliated Hospital of Luzhou Medical College, Luzhou 646000, China;
SUNXiaoyang, Email: sunxy0611@126.com
Export PDF Favorites Scan Get Citation

We applied Demons and accelerated Demons elastic registration algorithm in radiotherapy cone beam CT (CBCT) images, We provided software support for real-time understanding of organ changes during radiotherapy. We wrote a 3D CBCT image elastic registration program using Matlab software, and we tested and verified the images of two patients with cervical cancer 3D CBCT images for elastic registration, based on the classic Demons algorithm, minimum mean square error (MSE) decreased 59.7%, correlation coefficient (CC) increased 11.0%. While for the accelerated demons algorithm, MSE decreased 40.1%, CC increased 7.2%. The experimental verification with two methods of demons algorithm obtained the desired results, but the small difference appeared to be lack of precision, and the total registration time was a little long. All these problems need to be further improved for accuracy and reducing of time.

Citation: PANGHaowen, SUNXiaoyang. Application of Elastic Registration Based on Demons Algorithm in Cone Beam CT. Journal of Biomedical Engineering, 2014, 31(1): 103-106. doi: 10.7507/1001-5515.20140020 Copy

  • Previous Article

    Application of Semi-supervised Sparse Representation Classifier Based on Help Training in EEG Classification
  • Next Article

    CT Imaging Features and Their Correlation with Pathological Findings of Solid Pseudopapillary Tumor of Pancreas