• School of Medical Instruments and Food Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China;
SONGChengli, Email: csong@usst.edu.cn
Export PDF Favorites Scan Get Citation

Accurate measurements of voltage and current from electrosurgery are the basis of development of electrosurgery with feedback function. We, therefore, developed a parameter measurement system based on PC, with high voltage and current from electrosurgery being sensed with transformers, amplified, filtered, transformed into single-ended signals, and then into RMS signals. The root mean square (RMS) signals were transformed into digital signals through DAQ card and the data was processed in PC with Labview. The process included sampling, displaying and storage. The experiment results indicated that the measurement system could measure the output parameters from electrosurgery steadily and correctly so that the development of the system has been successful. It can be the basis of development of embedded parameters measurement system and can provide accurate feedback information for intellectual electrosurgery.

Citation: ZHOUYu, LIDianli, XUWendong, SONGChengli. A New Parameter Measurement System for Electrosurgery Output. Journal of Biomedical Engineering, 2014, 31(2): 421-425. doi: 10.7507/1001-5515.20140078 Copy

  • Previous Article

    An Approach for Segmentation of X-ray Angiographic Image Based on Region-growing and Structure Inferring
  • Next Article

    Experimental Study on Novel Hybrid Artificial Trachea Transplantation