This paper adopted UG8.0 to bulid the stent and blood vessel models. The models were then imported into the finite element analysis software ANSYS. The simulation results of ANSYS software showed that after endothelial stent implantation, the velocity of the blood was slow and the fluctuation of velocity was small, which meant the flow was relatively stable. When blood flowed through the endothelial stent, the pressure gradually became smaller, and the range of the pressure was not wide. The endothelial shear stress basically unchanged. In general, it can be concluded that the endothelial stents have little impact on the flow of blood and can fully realize its function.
Citation: SHIGengqiang, SONGXiaobing. Finite Element Analysis of Intravascular Stent Based on ANSYS Software. Journal of Biomedical Engineering, 2015, 32(5): 1004-1008. doi: 10.7507/1001-5515.20150178 Copy