ZHANGMei 1,2 , LINZhibo 1,2 , YUANPeng 1,2 , YAOZhifeng 1,2 , HUYueming 1,2
  • 1. Engineering Research Center for Precision Electronic Manufacturing Equipments of Ministry of Education, Guangzhou 510640, China;
  • 2. College of Automation Science and Engineering, South China University of Technology, Guangzhou 510640, China;
ZHANGMei, Email: zhangmei@scut.edu.cn
Export PDF Favorites Scan Get Citation

In view of the fact that medical inspection equipment sold in the domestic market is mainly imported from abroad and very expensive, we developed a full-automatic fluorescence analyzer in our center, presented in this paper. The present paper introduces the hardware architecture design of FPGA/DSP motion controlling card+PC+STM32 embedded micro processing unit, software system based on C# multi thread, design and implementation of double-unit communication in detail. By simplifying the hardware structure, selecting hardware legitimately and adopting control system software to object-oriented technology, we have improved the precision and velocity of the control system significantly. Finally, the performance test showed that the control system could meet the needs of automated fluorescence analyzer on the functionality, performance and cost.

Citation: ZHANGMei, LINZhibo, YUANPeng, YAOZhifeng, HUYueming. Integrated Development of Full-automatic Fluorescence Analyzer. Journal of Biomedical Engineering, 2015, 32(5): 1118-1124. doi: 10.7507/1001-5515.20150198 Copy

  • Previous Article

    Research on the Method of Blood Pressure Monitoring Based on Multiple Parameters of Pulse Wave
  • Next Article

    Segmentation Method for Liver Organ Based on Image Sequence Context