The present circuit was designed to apply to human tissue impedance tuning and matching device in ultrashort wave treatment equipment. In order to judge if the optimum status of circuit parameter between energy emitter circuit and accepter circuit is in well syntony, we designed a high frequency envelope detect circuit to coordinate with automatic adjust device of accepter circuit, which would achieve the function of human tissue impedance matching and tuning. Using the sampling coil to receive the signal of amplitude-modulated wave, we compared the voltage signal of envelope detect circuit with electric current of energy emitter circuit. The result of experimental study was that the signal, which was transformed by the envelope detect circuit, was stable and could be recognized by low speed Analog to Digital Converter (ADC) and was proportional to the electric current signal of energy emitter circuit. It could be concluded that the voltage, transformed by envelope detect circuit can mirror the real circuit state of syntony and realize the function of human tissue impedance collecting.

Citation: FANXu, WANGYunguang, CHENGHaiping, CHONGXiaochen. Design of High Frequency Signal Detecting Circuit of Human Body Impedance Used for Ultrashort Wave Diathermy Apparatus. Journal of Biomedical Engineering, 2016, 33(1): 89-95. doi: 10.7507/1001-5515.20160017 Copy

  • Previous Article

    Fast Implementation Method of Protein Spots Detection Based on CUDA
  • Next Article

    Design of Electrocardiogram Signal Generator Based on Typical Electrocardiogram Database