• Beijing Key Laboratory for Optoelectronic Measurement Technology, Beijing Information Science & Technology University, Beijing 100192, China;
GUOYangkuan, Email: guoyk2004@163.com
Export PDF Favorites Scan Get Citation

In order to solve the problem of the micro flow cell clogging, and to improve the reliability of the flow cytometry system, a new method was proposed for hydrodynamic self-cleaning system. By analyzing the flow cell focus principle, we considered that to obtain stable single cell flow, the stable pressure in the flow chamber must be ensured. Therefore, we established a diagnosis method of clogging by the pressure detecting, and designed a self-cleaning system. Then we built up corresponding experimental systems. Experiments and testing showed that the self-cleaning system could improve the flow and resolve the clogging problem.

Citation: WANGHaijun, GUOYangkuan, ZHULianqing, MENGXiaochen. Design of Flow Cytometry Self-cleaning System. Journal of Biomedical Engineering, 2016, 33(3): 570-574. doi: 10.7507/1001-5515.20160095 Copy

  • Previous Article

    Application of an Adaptive Inertia Weight Particle Swarm Algorithm in the Magnetic Resonance Bias Field Correction
  • Next Article

    Research Progress of Regulation of Hypoxia-mediated Signaling Pathways on Epithelial-Mesenchymal Transition of Tumor Cells