• School of Computer Science & Engineering, Northeastern University, Shenyang 110000, P. R. China;
ZHANG Shi, Email: zhangshi@mail.neu.edu.cn
Export PDF Favorites Scan Get Citation

The design of wall filter in ultrasonic microvascular imaging directly affects the resolution of blood flow imaging. We compared the traditional polynomial regression wall filter algorithm and two algorithms based on singular value decomposition (SVD), Full-SVD algorithm and RS-RSVD algorithm (random sampling based on random singular value decomposition) through experiments with simulated data and human renal entity data imaging experiments. The experimental results showed that the filtering effect of the traditional polynomial regression wall filter algorithm was limited, however, Full-SVD algorithm and RS-RSVD algorithm could better extract the micro blood flow signal from the tissue or noise signal. When RS-RSVD algorithm was randomly divided into 16 blocks, the signal-to-noise ratio was the same as that of Full-SVD algorithm, reduces the contrast-to-noise ratio by 2.05 dB, and reduces the execution time by 90.41%. RS-RSVD algorithm can improve the operation efficiency and is more conducive to the real-time imaging of high frame rate ultrasound microvessels.

Citation: WANG Baoyu, ZHANG Miao, LIU Ruilin, ZHANG Shi. Comparison of wall filter algorithms for ultrasonic microvascular imaging. Journal of Biomedical Engineering, 2022, 39(4): 740-748, 758. doi: 10.7507/1001-5515.202203032 Copy

  • Previous Article

    Medical nucleus image segmentation network based on convolution and attention mechanism
  • Next Article

    Mechanical and light-activated antibacterial properties of resin filled with Ag-TiO2 nanoparticles