Objective To explore the dyeing conditions of reticular fiber staining of liver puncture tissue stained by BenchMark automatic special staining instrument, evaluate the staining effect, and share the using experience.Methods A total of 30 cases of liver puncture specimens from April to May 2019 in West China Hospital of Sichuan University were selected. They were fixed with 4% neutral formaldehyde fix solution, treated with automatic dehydrator, embedded in paraffin, stained with adhesive slides, and baked in 65℃ oven for 30 min. Each specimen was sliced in duplicate, with a thickness of 5 μm. One was reticular fiber stained with BenchMark automatic special staining instrument under the dyeing condition according to the using experience of the laboratory, and the other was reticular fiber stained by manual Foot staining method according to the standard operating procedure of the laboratory. The staining effect under microscope was observed by a chief pathologist with extensive slide-reading experience.Results The success rate of instrumental method was 100.0% (30/30), and that of the manual method was 76.7% (23/30). The difference in success rate between the two methods was statistically significant (χ2=5.143, P=0.023).Conclusion The reticular fiber staining effect of BenchMark automatic special staining instrument is stable and reliable, and the success rate is higher than that of manual Foot staining.

Citation: CHEN Yu, YANG Mei, WANG Rong, LIAO Hongyan, MAO Zhigang, ZHENG Qin. Application of BenchMark automatic special staining instrument in reticular fiber staining of liver biopsy. West China Medical Journal, 2020, 35(9): 1090-1093. doi: 10.7507/1002-0179.201912082 Copy

  • Previous Article

    Construction of Z-score reference ranges of fetal ductus arteriosus inner diameter during mid- and late-pregnancy
  • Next Article

    The hundred top-cited articles in obstructive sleep apnea hypopnea syndrome: a bibliometric analysis