• Imaging Center, the First Affiliated Hospital, Xinjiang Medical University, Urumqi 830054;
LI Baiyan, Email: byl0318@163.com
Export PDF Favorites Scan Get Citation

Objective  To optimize image quality and radiation dose of infant chest digital radiography and to explore feasibility of reducing tube voltage and adjusting according to infant chest area.
Methods  0 to 3-year-old infants were randomly divided into two average groups of 0- and 1-3 year-old, and then each age group was randomly assigned to optimization and control groups in digital radiography. Measurement of radiation dose used dose area product (DAP). Mean DAP between groups was compared by using t test, and the image quality of optimization was compared by rank sum test.
Results  A total of 400 cases of 0 to 3-year-old infants were identified, and finally 391 cases of infants anteroposterior chest image were included, including 196 cases in the optimization group (0-years: n=91; 1-3 years: n=105) and 195 cases in the control group (0-years: n=103; 1-3 years: n=92). The results showed: there were significant differences in the mean DAP in 0-years, 1-3 years and total infants between the optimization group and the control group (all P value lt;0.05). The DAP of the optimization group was lower, and reduction of DAP was approximately 21.6% compared to the control subject. The Wilcoxon signed-rank test showed the difference of subjective evaluation of image quality was significantly different (P=0.000). High-quality image of the optimization group increased approximately 43.9% more than control subject.
Conclusion  Reducing tube voltage and adjusting according to infants chest area can not only reduce the radiation dose but also improve image quality in digital radiography.

Citation: GUO Hui,ZHANG Tieliang,NU Er,WU Haizhong,WU Jibin,LI Baiyan. Optimize Infants Chest Radiation Dose and Image Quality in DR: A Prospective Randomized Controlled Trial. Chinese Journal of Evidence-Based Medicine, 2011, 11(10): 1129-1132. doi: 10.7507/1672-2531.20110190 Copy

  • Previous Article

    Research on Influencing Factors and Circumcision Acceptance Willingness of Adult Males in Chongqing
  • Next Article

    Diagnostic Accuracy of Liquid-Based Cytology versus Conventional Cytology for Cervical Neoplasia: A Systematic Review of Randomized Controlled Trials