Abstract: Objective To summarize our operative experiences of cardiac reoperation after mechanical valve prosthesis replacement and investigate the causes of reoperation and the perioperative techniques and operation methods. Methods From January 2001 to December 2008, we performed reoperation on 105 patients (59 males and 46 females, aged 50.2±10.6 years old) who had undergone mechanical valve prosthesis replacement. Among the patients, there were 31 cases of mitral valvular replacement (+ tricuspid valvular plasticity), 38 cases of aortic valvular replacement (+ tricuspid valvular plasticity), 11 cases of Bentall procedure, 7 cases of mitral and aortic bivalvular replacement (+tricuspid valvular plasticity), 8 cases of tricuspid valvular replacement, 6 cases of repairing of prosthetic leakage, and 4 others cases. The time interval between two operations was 3 months to 18 years (46.3 ±31.9 months). Before reoperation, the cardiac function (NYHA) of the patients was class Ⅱ in 27 patients, class Ⅲ in 53 patients, and class Ⅳ in 25 patients. Results There were 6 hospital deaths with a mortality of 5.71%(6/105). All others recovered to NYHA class ⅠⅡ. The causes of mortality included 1 case of multiple organ failure, 1 case of low cardiac output after operation, 1 case of aortic pseudoaneurysm rupture, 1 case of severe infection due to brain complication and 2 cases of prosthetic valve endocarditis (PVE). The causes for cardiac reoperation after mechanical valve prosthesis replacement were 67 cases of prosthetic leakage (63.80%), 16 cases of PVE (15.23%), 14 cases of prosthetic thrombosis (13.33%) and 8 cases of other valvular anomalies. Followup was done for 11 to 107 months, which showed two cases late deaths of cardiac arrest and cerebral hemorrhage. Conclusion Patients who have received mechanical valve prosthesis replacement may undergo cardiac reoperation due to paravalvular prosthetic leakage, paravalvular endocarditis, and prosthetic thrombosis. The keys to a successful cardiac reoperation include appropriate preoperative preparations, operational timing, and suitable choosing of cardiopulmonary bypass and operational skills.
Citation: GUO Changfa,WANG Chunsheng,ZHAO Dong,et al .. Reoperation after Mechanical Valve Prosthesis Replacement: An Analysis of 105 Cases. Chinese Journal of Clinical Thoracic and Cardiovascular Surgery, 2010, 17(5): 361-364. doi: Copy