ObjectivesTo evaluate the reproducibility of Heidelberg retina tomograph (HRT) macular edema module(MEM) measuring the macular retinal thickness.MethodsSixty-two healthy volunteers (9-68 years old) were examined by HRT-II procedure. The retinal signal width (SW) at macula and fovea and macular edema index (E) were recorded for t-test, Pearson linear-correlation analysis. Intra-subject variation repeatedly measured was analyzed with coefficient of variation, 95% tolerance limits of change (TC), and intraclass coefficient of correlation (ICC). ResultsIn healthy individuals, retinal SW was (0.734±0.236) mm at macula,and (0.781±0.243) mm at fovea; macular E was (1.169±0.619). The coefficient of variation repeatedly measured: retinal SW was (8.7±68)%,retinal SW at the fovea was (8.5±6.7)%, and the average was (15.6±13.9)%; 95%TC of intra-subject sequential repeated measurement was 0.131 (8.9%) of retinal SW, 0.137 (10.5%)of fovea SW,and 0.198 (7.4%) of average E. ICC of one individual repeatedly measured by one operator was 0.950 of macular SW, 0.949 of fovea SW, and 0.898 of average edema index.ConclusionsHRT-II MEM is noninvasive, fast and highly reproducible, which provides a new technique to monitor the objective quantification of macular diseases related to retinal thickness. ( Chin J Ocul Fundus Dis, 2005,21:103-105)

Citation: LIANG Yuanbo,XU Liang,YU Qiurong,et al.. Reproducibility of Heidelberg retinal tomograph measuring the macular retinal thickness. Chinese Journal of Ocular Fundus Diseases, 2005, 21(2): 103-105. doi: Copy