Objective To compare the effect of electrode positions on multifocal visual evoked potential(mf-VEP). Methods Ten healthy individuals were tested with RETIscan multifocal VEP system 3.20, each individual was tested with four electrode positions: the active and reference electrode were placed 2.0 cm above and below the inion in Ch1;3.0 cm above and 4.5 cm below the inion in Ch2;2.0 cm on both right and left side of the inion in Ch3 and 4.0 cm on both right and left side of the inion in Ch4. The summed amplitudes of hemifield mf-VEP were analyzed and compared according to different electrode positions. Results The difference among summed amplitudes of the upper or lower hemifield mf-VEP recorded with Ch1、Ch2、Ch3 and Ch4 was statistically significant;and the difference between Ch2 and the other channels was statistically significant too. The difference among summed amplitudes of the horizontal sites whose amplitudes were small when recorded with Ch2 and summed amplitudes of the corresponding sites recorded with Ch3 and Ch4 were statistically significant;and the difference between Ch4 and the other channels was statistically significant too. Conclusions compared to Ch1、Ch3 and Ch4, Ch2 can make a better recording of mf-VEP. As for some sites, especially those along horizontal line, horizontal electrodes could improve the amplitudes of mf-VEP recorded with vertical electrodes,and Ch4 could do better to improve the mf-VEP recorded with Ch2 than Ch3. (Chin J Ocul Fundus Dis,2004,20:346-348)
Citation: SUN Chuanbin,DU Caifeng,CHEN Xiaoming. The effect of electrode positions on multifocal visual evoked potential. Chinese Journal of Ocular Fundus Diseases, 2004, 20(6): 346-348. doi: Copy