find Keyword "镜检法" 2 results
  • Comparison of Two Methods for Testing Urine Leukocyte

    目的:比较镜检法和尿液分析仪检测白细胞的差异。 方法: 用两种方法测定900例尿液标本中的白细胞,并对其结果进行比较分析。结果:两种方法检测结果有明显差异(Plt;005),以镜检法为标准,尿液分析仪检测白细胞假阳性率为169%,假阴性率为158%。结论: 检测尿液白细胞时镜检法和尿液分析仪尽量结合应用,以减少假阴性率和假阳性率,保证检验结果的准确性。

    Release date:2016-08-26 02:21 Export PDF Favorites Scan
  • Comparison between Instrument Method and Microscope Test Method in Measuring the Percentage of Immature Granulocytes of Peripheral Venous Blood in Patients with Leucocythemia

    【摘要】 目的 探讨仪器法和镜检法计数白血病幼稚粒细胞百分比的相关性和一致性。 方法 2009年6-9月对71例慢性粒细胞性白血病(慢粒)、亚急性粒细胞性白血病(M2)、急性早幼粒细胞性白血病(M3)及急性粒单核细胞性白血病(M4)白血病患者进行仪器法和镜检法计数周围静脉血幼稚粒细胞的百分比的检测,并进行比较分析。 结果 两种方法计数的慢粒、M2、M3及M4型共71例白血病患者的幼稚粒细胞的百分比比较,有统计学意义(t=6.404,Plt;0.01);但具有相关性(r=0.771,Plt;0.001)。且四种类型白血病中的每一种类型的白血病的两种方法的幼粒值也都具有相关性和不一致性(Plt;0.05)。 结论 SYSMEX XE-2100全自动血细胞分析仪对慢粒、M2、M3及M4的周围静脉血幼稚粒细胞识别能力欠佳,仍需采用镜检法进行检测。【Abstract】 Objective To investigate the correlation and consistency of the percentage of immature granulocytes of peripheral venous blood in patients with leucocythemia between the instrument method and microscope test method. Methods From June to September 2009, instrument method and microscope test method were used to measure the percentage of immature granulocytes of peripheral venous blood in patients with leucocythemia [chronic granulocytic leukemia (CGL), subacute granulocytic leukemia (M2), acute promyelocyte leukemia (M3), and acute myelomonocytic leukemia (M4)]. Results The difference in the percentage of immature granulocytes of the 71 samples between the 2 methods was significant (t=6.404,Plt;0.01), but still with correlation (r=0.771,Plt;0.001). Besides, there were also correlation and consistency of the percentage of immature granulocytes of the four types of leukemia between the two methods (Plt;0.05). Conclusion The identification ability of SYSMEX XE-2100 type automatic blood cell analyzer for measuring the percentage of immature granulocytes of peripheral venous blood in patients with CGL, M2, M3 and M4 is still weak. Currently, microscope test method still needs to be applied in the measurement.

    Release date:2016-09-08 09:51 Export PDF Favorites Scan