west china medical publishers
Author
  • Title
  • Author
  • Keyword
  • Abstract
Advance search
Advance search

Search

find Author "REN Jiaxuan" 2 results
  • Musculoskeletal multibody dynamics investigation for the different medial-lateral installation position of the femoral component in unicompartmental knee arthroplasty

    The surgical installation accuracy of the components in unicompartmental knee arthroplasty (UKA) is an important factor affecting the joint function and the implant life. Taking the ratio of the medial-lateral position of the femoral component relative to the tibial insert (a/A) as a parameter, and considering nine installation conditions of the femoral component, this study established the musculoskeletal multibody dynamics models of UKA to simulate the patients’ walking gait, and investigated the influences of the medial-lateral installation positions of the femoral component in UKA on the contact force, joint motion and ligament force of the knee joint. The results showed that, with the increase of a/A ratio, the medial contact force of the UKA implant was decreased and the lateral contact force of the cartilage was increased; the varus rotation, external rotation and posterior translation of the knee joint were increased; and the anterior cruciate ligament force, posterior cruciate ligament force and medial collateral ligament force were decreased. The medial-lateral installation positions of the femoral component in UKA had little effect on knee flexion-extension movement and lateral collateral ligament force. When the a/A ratio was less than or equalled to 0.375, the femoral component collided with the tibia. In order to prevent the overload on the medial implant and lateral cartilage, the excessive ligament force, and the collision between the femoral component and the tibia, it is suggested that the a/A ratio should be controlled within the range of 0.427−0.688 when the femoral component is installed in UKA. This study provides a reference for the accurate installation of the femoral component in UKA.

    Release date: Export PDF Favorites Scan
  • Effect of prosthetic joint line installation height errors on insert wear in unicompartmental knee arthroplasty

    The clinical performance and failure issues are significantly influenced by prosthetic malposition in unicompartmental knee arthroplasty (UKA). Uncertainty exists about the impact of the prosthetic joint line height in UKA on tibial insert wear. In this study, we combined the UKA musculoskeletal multibody dynamics model, finite element model and wear model to investigate the effects of seven joint line height cases of fixed UKA implant on postoperative insert contact mechanics, cumulative sliding distance, linear wear depth and volumetric wear. As the elevation of the joint line height in UKA, the medial contact force and the joint anterior-posterior translation during swing phase were increased, and further the maximum von Mises stress, contact stress, linear wear depth, cumulative sliding distance, and the volumetric wear also were increased. Furthermore, the wear area of the insert gradually shifted from the middle region to the rear. Compared to 0 mm joint line height, the maximum linear wear depth and volumetric wear were decreased by 7.9% and 6.8% at –2 mm joint line height, and by 23.7% and 20.6% at –6 mm joint line height, the maximum linear wear depth and volumetric wear increased by 10.7% and 5.9% at +2 mm joint line height, and by 24.1% and 35.7% at +6 mm joint line height, respectively. UKA prosthetic joint line installation errors can significantly affect the wear life of the polyethylene inserted articular surfaces. Therefore, it is conservatively recommended that clinicians limit intraoperative UKA joint line height errors to –2−+2 mm.

    Release date: Export PDF Favorites Scan
1 pages Previous 1 Next

Format

Content